TrialsgrambetaRCTs are beautiful

PARAMEDIC (LUCAS-2)

2015

Pre-hospital Randomised Assessment of a Mechanical Compression Device in Cardiac Arrest

In 4,471 out-of-hospital cardiac arrests, LUCAS-2 mechanical chest compression did not improve 30-day survival versus manual CPR, and worsened functional outcome at 3 months.

Result

Efficacy: 30-day survivaladj. OR 0.86 (95% CI 0.64-1.15)

Below 1 favours manual CPR, above 1 favours LUCAS-2 mechanical chest compression.

Event rates
Absolute difference
1 points

The absolute difference is derived from the event rates.

Also reported: CPC 1–2 at 3 months worse with LUCAS-2 (adj. OR 0.72).

Original publication
Subtopic
Mechanical CPR
Design
Cluster-randomised RCT, UK ambulance services
Relevance
Supporting: context rather than must-know.
Source
Perkins GD, et al. Lancet. 2015;385(9972):947-955.doi 10.1016/S0140-6736(14)61886-9

Discussion & critique

Mechanical CPR is not a routine replacement for high-quality manual CPR; reserve for transport, cath-lab and ECPR bridging.

Discussed in podcasts