PRAETORIAN

2020

Subcutaneous vs Transvenous Defibrillator Therapy

In 849 patients needing a defibrillator but no pacing, a subcutaneous ICD was non-inferior to a transvenous ICD for device complications and inappropriate shocks.

Result

Efficacy: Device complications + inappropriate shocksHR 0.99 (95% CI 0.71-1.39) · P = 0.01

Below 1 favours subcutaneous ICD, above 1 favours a transvenous ICD.

Event rates
Absolute difference
0.6 points

The absolute difference is derived from the event rates.

Also reported: Non-inferior; P=0.95 superiority.

Original publication
Subtopic
ICD
Design
RCT, non-inferiority
Relevance
Important: one of several pillars.
Source
Knops RE, et al. N Engl J Med. 2020;383(6):526-536.doi 10.1056/NEJMoa1915932

Discussion & critique

Trades lead complications for more inappropriate shocks and no pacing or anti-tachycardia therapy. Suits patients with no pacing or resynchronisation indication.