OptimPacing

2026

OptimPacing: Protection of Cardiac Function With Left Bundle Branch Pacing in Patients With Atrioventricular Block

In adults with second-degree or complete atrioventricular block, or ongoing atrial fibrillation, with a slow heart rate and related symptoms, left bundle branch pacing reduced death, heart failure hospitalisation or pacing-induced cardiomyopathy compared with right ventricular pacing.

Risultato

All-cause death, heart failure hospitalisation or pacing-induced cardiomyopathy (assessed at 6 to 36 months)
Differenza assoluta
10,7 punti
Stima dell’effetto

59% lower risk59

La differenza assoluta è derivata dai tassi di eventi.

Riportato inoltre: At four years, No patient developed pacing-induced cardiomyopathy with left bundle branch pacing vs 11.7% with right ventricular pacing. Heart failure hospitalisation occurred in 1.5% vs 2.9%.

Pubblicazione originale
Sottotema
Conduction system pacing
Disegno
Double-blind randomised trial (First Affiliated Hospital of Nanjing Medical University; the programme calls it international and multicentre) of left bundle branch pacing versus right ventricular pacing in AV block with LVEF above 35%
Arruolati
685 pazienti
Fonte
Qian Z, Jiang L, Hou X, et al. N Engl J Med. 2026; published online 8 November.doi 10.1056/nejmoa2612972

Discussione e critica

In this randomised trial, left bundle branch pacing roughly halved a composite of death, heart failure hospitalisation and pacing-induced cardiomyopathy, and much of the gap came from cardiomyopathy, which did not occur at all with left bundle branch pacing. The trial was stopped early at a planned interim analysis after a median 3-year follow-up, which the investigators say may overstate the benefit and make the estimate less precise. The newsroom report gives no hazard ratio, confidence interval or P value, and it says only that the event committee was blinded, not patients or operators. Most participants had a normal ejection fraction and every left bundle branch procedure used the same lead type, so the results may not carry over to reduced heart function or to other pacing tools.