DeepRhythmAI
2025
Artificial Intelligence for Direct-to-Physician Reporting of Ambulatory Electrocardiography
In 14,606 ambulatory ECG recordings annotated beat by beat, AI interpretation reported directly to the physician improved sensitivity for critical arrhythmias against cardiologist consensus panels compared with certified ECG technicians.
Result
Efficacy: Sensitivity for critical arrhythmias versus cardiologist consensus-panel annotationrelative risk of a missed diagnosis 14.1 (95% CI 10.4-19.0)
Below 1 favours certified ECG technicians, above 1 favours AI interpretation reported directly to the physician.
The estimate refers to a different reading of the endpoint than the rates.
Also reported: False positives 12 vs 5 per 1,000 patient-days.
- Subtopic
- Atrial fibrillation
- Design
- Non-randomised diagnostic-accuracy comparison against 17 cardiologist consensus panels
- Relevance
- Supporting: context rather than must-know.
- Source
- Johnson LS, et al. Nat Med. 2025;31(3):925-931.doi 10.1038/s41591-025-03516-x
Discussion & critique
The largest head-to-head of AI against human technicians for ambulatory ECG, supporting AI-first reporting with physician review. It is an accuracy study, not an outcome trial: the reference standard was a cardiologist-adjudicated random sample of 5,235 events rather than every recording, the comparison was not randomised, and the price of the very high sensitivity was a higher false-positive burden that physicians must absorb.